Inicio  /  Applied Sciences  /  Vol: 9 Par: 16 (2019)  /  Artículo
ARTÍCULO
TITULO

Calibration Method for Angular Positioning Deviation of a High-Precision Rotary Table Based on the Laser Tracer Multi-Station Measurement System

Hongfang Chen    
Bo Jiang    
Hu Lin    
Shuang Zhang    
Zhaoyao Shi    
Huixu Song and Yanqiang Sun    

Resumen

This paper proposes a calibration method for angular positioning deviation of a high-precision rotary table based on the laser tracer multi-station measurement system. The algorithm error of the calibration method for angular positioning deviation of a high-precision rotary table based on the laser tracer multi-station measurement system was mainly discussed. During the experiments, the laser tracer was fixed on the work surface of the rotary table, and the rotary was fixed on the work surface of the coordinate measurement machine (CMM). The rotary table was rotated with the same angular interval. In this case, an optimization method for calculating the coordinates of a laser tracer station by using Levenberg?Marquardt algorithm and singular value decomposition transform was proposed. Then, the angular positioning deviation of the rotary table was calibrated by an established geometric relationship model between the coordinates of laser tracer stations and the rotation angle of the rotary table. The angular positioning deviation of the high-precision rotary table was as low as ±0.9?, and the error of the calibration method was ±0.4?. The experimental results proved the feasibility of the proposed calibration method. The calibration method proposed in this paper is suitable for the case that the rotary table is not linked with the CMM, especially for large high-precision rotary tables.