A High-Resolution Mass Spectrometer for the Experimental Study of the Gas Composition in Planetary Environments: First Laboratory Results

Illia Zymak    
Ján ?abka    
Miroslav Polá?ek    
Arnaud Sanderink    
Jean-Pierre Lebreton    
Bertrand Gaubicher    
Barnabé Cherville    
Anna Zymaková and Christelle Briois    

Resumen

A new laboratory OrbitrapTM cell-based mass spectrometer, OLYMPIA (Orbitrap anaLYseur MultiPle IonisAtion), without a C-trap module, has been developed and constructed. The first operation of the OrbitrapTM cell-based device with the continuous ion source and without the C-trap module is reported. OLYMPIA is being developed and used as a workbench platform to test and develop technologies for the next generation of spaceborne mass spectrometers and as a laboratory instrument to perform high-resolution studies of space-relevant chemical processes. This instrument has been used to measure the quantitative composition of CO/N2/C2H4 mixtures of the same nominal mass using an electron ionization ion source. The relative abundance of ions has been measured using a short acquisition time (up to 250 ms) with a precision of better than 10% (for most abundant ions) and a mass resolution of 30,000?50,000 (full width at half maximum) over the mass range of m/z 28?86. The achieved mass accuracy of measurements is better than 20 ppm. This performance level is sufficient to resolve and identify the CO/N2/C2H4 components of the mixtures. The dynamic range and relative ion abundance measurements have been evaluated using a reference normal isotopic distribution of krypton gas. The measurement accuracy is about 10% for the 4 most abundant isotopes; 6 isotopes are detectable.