Resumen
A method for volumetric Tensor Velocity Imaging employing row-column (RC) addressed array with low computational complexity is investigated in simulations. An interleaved and non-interleaved sliding aperture sequence with 11 rows and 11 columns emissions by a 62 + 62 RC addressed array was used. The 3D velocities were estimated by a transverse oscillation (TO) cross-correlation estimator. Parabolic profiles at six different orientations corresponding to combinations of 0, 45 degrees azimuth angles and 90, 75, 60 beam-to-flow angles were investigated with 5 kHz pulse repetition frequencies. The Field II simulations were performed at a depth of 30 mm with peak velocity of 0.3 m/s. Across all vessel orientations, the relative mean bias varied from 2.3% to -14.26%, and the relative standard deviation varied from 0.43% to 5.5%. The best and worst performance was found at beam to flow angles of 90 degrees with 0 degrees rotation angle and 60 degrees beam-to-flow angle with 45 degrees rotation angle respectively. Due to the low channel count of the RC array and the low computational complexity, real-time implementation is feasible on conventional ultrasound systems.