Inicio  /  Algorithms  /  Vol: 16 Par: 6 (2023)  /  Artículo
ARTÍCULO
TITULO

Clinical Validation of a New Enhanced Stent Imaging Method

Chadi Ghafari    
Khalil Houissa    
Jo Dens    
Claudiu Ungureanu    
Peter Kayaert    
Cyril Constant and Stéphane Carlier    

Resumen

(1) Background: Stent underexpansion is the main cause of stent thrombosis and restenosis. Coronary angiography has limitations in the assessment of stent expansion. Enhanced stent imaging (ESI) methods allow a detailed visualization of stent deployment. We qualitatively compare image results from two ESI system vendors (StentBoost? (SB) and CAAS StentEnhancer? (SE)) and report quantitative results of deployed stents diameters by quantitative coronary angiography (QCA) and by SE. (2) Methods: The ESI systems from SB and SE were compared and graded by two blinded observers for different characteristics: 1 visualization of the proximal and distal edges of the stents; 2 visualization of the stent struts; 3 presence of underexpansion and 4 calcifications. Stent diameters were quantitatively measured using dedicated QCA and SE software and compared to chart diameters according to the pressure of implantation. (3) Results: A total of 249 ESI sequences were qualitatively compared. Inter-observer variability was noted for strut visibility and total scores. Inter-observer agreement was found for the assessment of proximal stent edge and stent underexpansion. The predicted chart diameters were 0.31 ± 0.30 mm larger than SE diameters (p < 0.05). Stent diameters by SE after post-dilatation were 0.47 ± 0.31 mm smaller than the post-dilation balloon diameter (p < 0.05). SE-derived diameters significantly differed from QCA; by Bland?Altman analysis the bias was -0.37 ± 0.42 mm (p < 0.001). (4) Conclusions: SE provides an enhanced visualization and allows precise quantitative assessment of stent expansion without the limitations of QCA when overlapping coronary side branches are present.

 Artículos similares

       
 
Ana Teresa Gabriel, Cláudia Quaresma and Pedro Vieira    
Back pain is regularly associated with biomechanical changes in the spine. The traditional methods to assess spine biomechanics use ionising radiation. Vertebral Metrics (VM) is a non-invasive instrument developed by the authors in previous research that... ver más
Revista: Algorithms

 
Gianluca Bochicchio, Luca Ferrari, Alberto Bottari, Francesco Lucertini, Valentina Cavedon, Chiara Milanese and Silvia Pogliaghi    
Force?velocity profiling (F-v) evaluates muscle function through the identification of maximum force (F0), velocity (V0), power (Pmax) and optimal velocity (Vopt). The purpose of this study was to investigate the validity and feasibility of loaded 5 Sit-... ver más
Revista: Applied Sciences

 
Massimiliano Greco, Pier Francesco Caruso, Sofia Spano, Gianluigi Citterio, Antonio Desai, Alberto Molteni, Romina Aceto, Elena Costantini, Antonio Voza and Maurizio Cecconi    
Background: Sepsis is one of the major causes of in-hospital death, and is frequent in patients presenting to the emergency department (ED). Early identification of high-risk septic patients is critical. Machine learning (ML) techniques have been propose... ver más
Revista: Algorithms

 
Wentao Lv, Fan Li, Shijie Luo and Jie Xiang    
Autism spectrum disorder (ASD) is a complex neurodevelopmental disorder that can reduce quality of life and burden families. However, there is a lack of objectivity in clinical diagnosis, so it is very important to develop a method for early and accurate... ver más
Revista: Algorithms

 
Xingchen Xu, Xingguang Geng, Zhixing Gao, Hao Yang, Zhiwei Dai and Haiying Zhang    
The accurate localization of S1 and S2 is essential for heart sound segmentation and classification. However, current direct heart sound segmentation algorithms have poor noise immunity and low accuracy. Therefore, this paper proposes a new optimal heart... ver más
Revista: Applied Sciences