Redirigiendo al acceso original de articulo en 20 segundos...
Inicio  /  Applied Sciences  /  Vol: 13 Par: 1 (2023)  /  Artículo
ARTÍCULO
TITULO

Characterization of Ultra-High-Dose Rate Electron Beams with ElectronFlash Linac

Lucia Giuliano    
Gaia Franciosini    
Luigi Palumbo    
Lilia Aggar    
Marie Dutreix    
Luigi Faillace    
Vincent Favaudon    
Giuseppe Felici    
Federica Galante    
Andrea Mostacci    
Mauro Migliorati    
Matteo Pacitti    
Annalisa Patriarca and Sophie Heinrich    

Resumen

Purpose: The electron linac ElectronFlash installed at Institut Curie (Orsay, France) is entirely dedicated to FLASH irradiation for radiobiological and pre-clinical studies. The system was designed to deliver an ultra-high-dose rate per pulse (UHDR) (above 106" role="presentation" style="position: relative;">106106 10 6 Gy/s) and a very high average dose rate at different energies and pulse durations. A campaign of tests and measurements was performed to obtain a full reliable characterizations of the electron beam and of the delivered dose, which are necessary to the radiobiological experiments. Methods: A Faraday cup was used to measure the electron charges in a single RF pulse. The percentage depth dose (PDD) and the transverse dose profiles, at the energies of 5 MeV and 7 MeV, were evaluated employing Gafchromic films EBT-XD for two Poly-methylmethacrylate (PMMA) applicators with irradiation sizes of 30 mm and 120 mm, normally used for in vivo and in vitro experiments, respectively. The results were compared with Monte Carlo (MC) simulations. Results: The measurements were performed during a period of a few months in which the experimental set up was adapted and tuned in order to characterize the electron beam parameters and the values of delivered doses before the radiobiological experiments. The measurements showed that the dose parameters, obtained at the energy of 5 MeV and 7 MeV with different applicators, fulfill the FLASH regime, with a maximum value of an average dose rate of 4750 Gy/s, a maximum dose per pulse of 19 Gy and an instantaneous dose rate up to 4.75 ×106" role="presentation" style="position: relative;">×106×106 × 10 6 Gy/s. By means of the PMMA applicators, a very good flatness of the dose profiles was obtained at the cost of a reduced total current. The flatness of the large field is reliable and reproducible in radiobiological experiments. The measured PDD and dose profiles are in good agreement with Monte Carlo simulations with more than 95%" role="presentation" style="position: relative;">95%95% 95 % of the gamma-index under the thresholds of 3 mm/3%" role="presentation" style="position: relative;">3%3% 3 % . Conclusions: The results show that the system can provide UHDR pulses totally satisfying the FLASH requirements with very good performances in terms of beam profile flatness for any size of the fields. The monitoring of electron beams and the measurement of the dose parameters played an important role in the in vivo and in vitro irradiation experiments performed at the Institut Curie laboratory.

Palabras claves

 Artículos similares

       
 
Juyu Wang, Shenyi Zhang, Guohong Shen, Ying Sun, Binquan Zhang, Zheng Chang, Chunqin Wang, Donghui Hou and Zhe Yang    
Based on orbit detection data acquired by a positive channel Metal Oxide Semiconductor (PMOS) dose detectors on FY4-A (GEO), BD3-M15 (MEO), and YH1-01A (LEO) between November 2018 and November 2022, investigations reveal variations in total dose and the ... ver más
Revista: Aerospace

 
Fahad A. Alharthi, Riyadh H. Alshammari and Imran Hasan    
The elimination of toxic heavy metal ions from wastewater has been found to be of great importance in human as well marine animal wellbeing. Among various heavy metals, Cr (VI) has been found to be one of the highly toxic and carcinogenic heavy metals wh... ver más
Revista: Water

 
Izabela M. Domanska, Ramona Figat, Aldona Zalewska, Krystyna Ciesla, Sebastian Kowalczyk, Karolina Kedra and Marcin Sobczak    
The effect of ionizing radiation (?-rays and electron beam) on anticancer drug delivery systems (DDSs) properties was evaluated concerning potential sterilization. For this purpose, paclitaxel (PTX)-loaded nanoparticles were obtained using a biodegradabl... ver más
Revista: Applied Sciences

 
Manasik M. Nour, Maha A. Tony and Hossam A. Nabwey    
The ever-increasing technological advancement and industrialization are leading to a massive discharge of hazardous waste into the aquatic environment, calling on scientists and researchers to introduce environmentally benign solutions. In this regard, t... ver más
Revista: Applied Sciences

 
Min Jeong Hong, Dae Yeon Kim, Yeong Deuk Jo, Hong-Il Choi, Joon-Woo Ahn, Soon-Jae Kwon, Sang Hoon Kim, Yong Weon Seo and Jin-Baek Kim    
Gamma rays as a type of ionizing radiation constitute a physical mutagen that induces mutations and could be effectively used in plant breeding. To compare the effects of gamma and ionizing irradiation according to exposure time in common wheat (Keumgang... ver más
Revista: Applied Sciences