This study presents the feasibility of a dual-mode high-pass birdcage RF coil to acquire MR images at both 1H and 23Na frequencies at ultra-high-field MR scanner, 7 T. A dual-mode circuit (DMC) in the dual-mode birdcage (DMBC) RF coil operates at two frequencies, addressing the limitations of sensitivity reduction and isolation between two frequencies as in traditional dual-tuned RF coil. Finite-difference time-domain (FDTD) based electromagnetic (EM) simulations were performed to verify the RF coil at each frequency on the three-dimensional human head model. The DMBC RF coil resonated at proton (1H) and sodium (23Na) frequencies, and also single-tuned high-pass birdcage RF coils were constructed for both 1H and 23Na frequencies. The bench test performance of the RF coils was evaluated using network analysis parameters, including the measurement of scattering parameters (S-parameters) and quality factors (Q-factors). Q-factor of the DMBC coil at 1H port was 10.2% lower than that of 1H single-tuned birdcage (STBC) coil, with a modest SNR reduction of 6.5%. Similarly, the Q-factor for the DMBC coil at 23Na port was 12.3% less than that of 23Na STBC coil, and the SNR showed a minimal reduction of 5.4%. Utilizing the DMBC coil, promising 1H and 23Na MR images were acquired compared to those by using STBC coils. In conclusion, deploying a DMBC 1H/23Na coil has been demonstrated to overcome traditional constraints associated with dual-tuned RF coils, achieving this with only nominal signal attenuation across both nuclei operational frequencies.