In-Vitro Investigation of Fatigue and Fracture Behavior of Transmucosal versus Submerged Bone Level Implants Used in Fixed Prosthesis

Saverio Cosola    
Paolo Toti    
Enrico Babetto    
Ugo Covani    
Miguel Peñarrocha-Diago and David Peñarrocha-Oltra    

Resumen

Background: The present in vitro study aimed to investigate the fatigue performance of different dental fixtures in two different emergence profiles. Biological failures are frequently reported because the problem canonly be solved by replacing a failing implant with a new one. Clinicians addressed minor mechanical failures, such as bending, loosening or the fracture of screws, abutment, or the entire prosthesis, by simply replacing or fixing them. Methods: Transmucosal and submerged bone-level dental implants underwent fatigue strength tests (statical and dynamical performance) by a standardized test: UNI EN ISO 14801:2016. Two types of emergence profiles (Premium sub-crestal straight implant with a cylindrical-shaped coronal emergence or Prama one-piece cylindrical-shape implant with transmucosal convergent neck and hyperbolic geometry) were tested, and dynamic fatigue were run to failure. Data was analyzed by a suitable statistical tool. Results: The Wöhler curve of 0.38 cm Premium group c2, appeared to be significantly different from that of the 0.38 cm Prama group c3 (nonparametric one-way ANOVA ?2 = 6; degree of freedom = 1; probability = 0.0043) but not from that of the 0.33 cm Premium group c1 (nonparametric one-way ANOVA ?2 = 0.62; degree of freedom = 1; probability = 0.4328). Fatigue performance of configuration 2 was one and a half times better than that of configuration 3. Group c3 had a better ultimate failure load (421.6 ± 12.5 N) than the other two settings i.e., c1 (324.5 ± 5.5 N) and c2 (396.3 ± 5.6) reaching almost a nonsignificant level. Conclusions: It was observed that a transmucosal implant design could provide the highest resistance to static fracture. On the other hand, an equicrestal implant design could increase dynamic endurance.

 Artículos similares

       
 
Roberta Pecoraro, Santi Concetto Pavone, Elena Maria Scalisi, Carmen Sica, Sara Ignoto, Martina Contino, Antonio Salvaggio, Dimitra Marmara, Gino Sorbello, Loreto Di Donato and Maria Violetta Brundo    
Recently, an increasing use of wireless internet technologies has been demonstrated. The devices which use these technologies emit in new spectral regions an electromagnetic radiation (EMF) which could interact with the male reproductive system. The aim ... ver más

 
Hany Ezzat Khalil, Hairul-Islam Mohamed Ibrahim, Kareem Ahmed El-Fass, Sabah H. Akrawi and Mohamed A. Morsy    
Liver inflammation is associated with an increased risk of liver fibrosis that substantially progresses to cirrhosis. Recently, usage of the herbal supplement has been increased because of its emerging role to dominate oxidative stress in hepatic injury.... ver más
Revista: Applied Sciences

 
Roberto De Santis, Flavia Iaculli, Vincenzo Lodato, Vito Gallicchio, Michele Simeone, Gianrico Spagnuolo and Carlo Rengo    
Background: Enhancement of the temperature of sodium hypochlorite (NaOCl) solution would increase its cleaning potential and decontamination of the root canal system. Therefore, the aim of the present in vitro investigation was to compare the efficacy of... ver más
Revista: Applied Sciences

 
Jong-ryul Choi and Juyoung Park    
Techniques that increase the permeability of the cell membrane and transfer drugs or genes to cells have been actively developed as effective therapeutic modalities. Also, in line with the development of these drug delivery techniques, the establishment ... ver más
Revista: Applied Sciences

 
José Ariévilo Gurgel Rodrigues, Ana Luíza Gomes Quinderé, Norma Maria Barros Benevides (Author)     Pág. e49082
The structural complexity of the agaran type-sulfated polysaccharides (SPs) found in Acanthophora muscoides limits its investigation as anticoagulant alternative to heparin which induces clot complications. This study was extended to evaluate the propert... ver más